Series 8035XA Peak Power Sensors
2-16
Manual 21568, Rev. F, March 2008
Examples:
OUTPUT 713;TR3
! Last measured value will be returned
OUTPUT 713;TR2
! Refill averaging buffer before measurement display
OUTPUT 713;TR1
! Wait for trigger before returning measurement
OUTPUT 713;TR0
! Measure, but no display
TR0
The meter will measure power, but the display of measured data will be suppressed and the
GPIB bus will not be updated with measurement data.
TR1
This mode will wait until the sensor triggers before returning a measurement. The
measurement returned will be after the application of any averaging. The display for the
channel will follow the TR1 mode. When TR1 is received over the bus,
NO TRIG
will be
displayed until the sensor has triggered and measurement data is available.
TR2
This mode will wait until enough measurements are made to completely refill the averaging
buffer. The measurement returned will be the average of all measurements in the buffer.
MEAS*
will display with one of the asterisk lines rotating for each measurement) while the
averaging buffer is being filled.
TR3
The last measured value will be returned. This mode will not wait for the peak sensor to trigger.
2.4.2
Setting Delays
When the sensor is configured for internal triggering, the delay from trigger to measurement
sample must be set. The valid range of delays is -20 ns to 100 ms, expressed in a floating point
number. The smallest delay increment is 0.5 ns. Setting delays in CW trigger mode are invalid
and ignored.
Examples:
OUTPUT 713;PEAK A DELAY 1.20E-6
! Configure sensor A for a delay of 1.20
µ
s
OUTPUT 713;PEAK B DELAY 33.5E-9
! Configure sensor B for a delay of 33.5 ns
The offset command adds a known offset to the trigger delay value. The actual value of delay
would be the DELAY set plus the OFFSET set. The default value of offset is 0. The valid range
of offset is -20 ns to 100 ms, expressed in a floating point number.
Example:
OUTPUT 713;PEAK A OFFSET 1.00E-6
! Configure sensor A for a delay offset of 1.00
µ
s
Содержание 80350A
Страница 4: ......
Страница 10: ...Series 8035XA Peak Power Sensors vi Manual 21568 Rev F March 2008 ...
Страница 13: ...Manual 21568 Rev F March 2008 ...
Страница 16: ...Series 8035XA Peak Power Sensors xii Manual 21568 Rev F March 2008 ...
Страница 17: ...Manual 21568 Rev F March 2008 ...
Страница 18: ...Series 8035XA Peak Power Sensors xiv Manual 21568 Rev F March 2008 ...
Страница 24: ...Series 8035XA Peak Power Sensors 1 6 Manual 21568 Rev F March 2008 ...
Страница 49: ...Figure 3 5 Digital Timing Diagram INT EXT Trig Mode Theory of Operation Manual 21568 Rev F March 2008 ...
Страница 90: ...Series 8035XA Peak Power Sensors 6 22 Manual 21568 Rev F March 2008 ...
Страница 92: ......
Страница 93: ......
Страница 94: ......
Страница 95: ......
Страница 96: ......
Страница 97: ......
Страница 98: ......
Страница 99: ......
Страница 100: ......
Страница 101: ......
Страница 102: ......
Страница 103: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......
Страница 107: ......