11. Frequently asked questions [FAQ]
Problem:
The logger cannot be stopped via keys.
Possible reason:
“Logger stop via buttons disabled” has been activated at the configu-
ration of the logger.
Remedy:
Deactivate this function by means of GSOFT
40K (at tab „settings“).
Problem:
The logger cannot be stopped both via keys and GSOFT 40K.
Possible reason:
“Do not stop active logger” has been activated at the configuration of
the logger.
Remedy:
Deactivate this function by means of EASYBus-configurator.
Problem:
The alarm settings cannot be changed.
Possible reason:
The logger has still saved data.
Remedy:
Stop logger as recommended by the software (delete data).
Problem:
A error message is displayed.
Error message
Possible reason
procedure / remedy
measuring range or
display range ex-
ceeded
Measuring range or display
range has been exceeded.
Operate logger within its specifica-
tions.
measuring range or
display range un-
der-run
Measuring range or display
range is under-run.
Operate logger within its specifica-
tions.
system error
Recording error
Reset the error via GSOFT 40K.
If the error message is displayed
furthermore send the logger to the
manufacturer for repair.
System error
Send the logger to the manufactur-
er for repair.