INTRODUCTION
The
HD402…
series is composed of pressure transmitters and pressure ON/OFF relay
switches. The instruments measure relative pressure with respect to atmosphere or
differential pressure in the range from 50 Pa to 200 kPa (depending on model) by us-
ing a silicon piezoresistive sensor with high accuracy and temperature compensation,
which allows excellent linearity, repeatability and stability over the time.
Thanks to the particular sensor used, the instruments are insensitive to the mounting
orientation and position. Moreover, the high stability of the sensor over the time and
in comparison to the changes in temperature allows eliminating the operations of
maintenance typically required to compensate for the aging and the deviation of the
sensor zero. The auto-zeroing feature in the HD402TR1L low range model allows sta-
ble measurements over the time without the need to recalibrate.
Available outputs, depending on the model:
•
RS485 Modbus–RTU digital output (HD402ST…);
•
0...10 V voltage analog output (HD402T…);
•
0...20 mA / 4...20 mA active current analog output (HD402T…);
•
2-wire (current loop) 4...20 mA analog output (HD402AT…);
•
ON/OFF relay switch (HD402TR…L).
Different units of measurement can be chosen for each model and, in the models with
analog output, it is possible to choose the full scale (f.s.) value for the output (high,
intermediate or low range) and set the unipolar (0...+f.s.) or bipolar (-f.s....+f.s.)
measuring range.
Versions with optional 4-digit LCD display (option
L
) are available, which allow the
measured parameters to be displayed. The pressure relay switches always include the
display.
The configuration can be made by connecting the serial port of the instrument to the
PC or by means of the quick configuration dip switches on the circuit board. The ver-
sions with relay switch output are equipped with two internal buttons for the configura-
tion via display.
The instruments are factory calibrated and ready for use.
Models
HD402
“” = without LCD (except HD402TR…)
L
= with LCD
1
= range ± 250 Pa / 25 mmH
2
O / 1 inchH
2
O / 2,5 mbar
2
= range ± 1000 Pa / 100 mmH
2
O / 4 inchH
2
O / 10 mbar
3
= range ± 10 kPa / 50 mmHg / 1,5 PSI / 100 mbar
4
= range ± 100 kPa / 500 mmHg / 15 PSI / 1000 mbar
5
= range ± 200 kPa / 1000 mmHg / 30 PSI / 2000 mbar
T
= 0…10 V and 0/4…20 mA active analog output
AT
= 2-wire 4…20 mA analog output
ST
= RS485 Modbus-RTU digital output
TR
= ON/OFF relay switch output