is set to OUTS.
•
Hysteresis
(HYST): Value of the hysteresis for above (RISE) and below (FALL)
threshold alarm operating modes. The parameter does appear if the alarm operat-
ing mode is set to OUTS.
•
Alarm activation delay
(T1): Value in seconds of the time delay for generating
the alarm. The alarm is generated only if the measurement exceeds the thresh-
old for more than the set time.
•
Alarm deactivation delay
(T2): Value in seconds of the time delay for deac-
tivating the alarm. The alarm is de-activated only after the set time has elapsed
from the disappearance of the alarm condition.
Note: the EDGE, THR1, THR2, HYST, T1 and T2 parameters do not appear if the alarm
is set to OFF.
Changing non-numerical parameters:
•
Select the parameter by using the button
B1
.
•
Short press the button
B2
to change the setting.
•
Long press the button
B1
to move to the next parameter.
Changing numerical parameters:
•
Select the parameter by using the button
B1
.
•
Short press the button
B2
to change the sign.
•
Long press the button
B2
to select the first digit.
•
Short press the button
B2
to change the selected digit.
•
Long press the button
B2
to select the next digit.
•
Repeat the above two steps until all the digits are set.
•
Long press the button
B1
to move to the next parameter.
While a parameter value is displayed, a short press of the button
B1
will briefly show
the name of the currently selected parameter.
4.4
E
RROR MESSAGES ON DISPLAY
Undr
: The measured value is less than the minimum measurable value.
OvEr
: The measured value exceeds the maximum measurable value.
CAL Error
: It appears at the end of the zero calibration if the maximum offset value
possible to be corrected is exceeded.