Bedienungsanleitung
As of 04/20/2010
Seite 15/16
Hint
Before updating the firmware there must be made sure that only one PAB device (this
Expansion Box) is connected directly to the ProfiLux. Other PAB devices at the
ProfiLux during the update are
not
allowed!
During the firmware update the sensors are not read, measurement values are not available. Interface
outputs (pumps, illuminations) aren’t accessed anymore. Therefore all accessed devices (like pumps,
valves, switching plugs, dosing pumps …) should be disabled or disconnected. Sensors or illuminations are
generally noncritical and can remain connected.
7
Technical data
Input voltage
12 VDC
Power consumption
200 mA
pH-measurement
BNC input for pH-sensors, accuracy 0.1 pH, measurement range
3.0 pH to 10.5 pH
Temperature measurement
Mini-DIN socket for delivered temperature sensor, accuracy 0.1 °C,
measurement range 11.5 °C to 38 °C
Conductivity measurement
BNC input for conductivity sensor, in fresh water accuracy 1 µS,
measurement range 0 µs to 2000 µS, in salt water accuracy 0.1 mS,
0 mS to 100 mS
Redox measurement
BNC input for Redox sensor, accuracy 1 mV, measurement range
-1000 mV to 1000 mV
Level inputs
Mini-DIN socket for 2 level inputs
Control dimmable lamps
2 modular jacks a‘ 2 dimming channels with analog output each 1-
10 V for light intensity control and shut off
Dimensions
H x B x T = 58 mm x 240 mm x 198 mm
Module slots
5 (4 for all module card types, one only for analog measurement
cards)