background image

4

Signet 9900 Conductivity Transmitter 

Installation

Panel  Cutout

92  x 92 mm

(+ 0.8, - 0 mm)

3.6 x 3.6 in.

(+0.031, -0 in.)

Panel  Cutout

92  x 92 mm

(+ 0.8, - 0 mm)

3.6 x 3.6 in.

(+0.031, -0 in.)

Panel  Cutout

92  x 92 mm

(+ 0.8, - 0 mm)

3.6 x 3.6 in.

(+0.031, -0 in.)

Panel  Cutout

92  x 92 mm

(+ 0.8, - 0 mm)

3.6 x 3.6 in.

(+0.031, -0 in.)

minimum

clearance

25 mm

(1 in.)

panel

Quick-clip

gasket on
front side
of panel

terminal
blocks

mounting

bracket

DC Power

Loop V

oltage

3-9900.394

Cond/Res Module

Do not mount in direct sunlight.

System Start-up: Step 1
Prepare the transmitter installation location.
If the back of the transmitter is diffi cult to access when installed, wire the 
removable terminal blocks fi rst, then complete the installation.

Next step: Wiring (Step 2).

Panel Mount Installation

Tools and Equipment Required

 

Fine-tooth 

fi

 le

 

¼ DIN punch or jigsaw suitable for cutting panel opening to within 1 mm (0.04 in) 
tolerance.

 

¼ DIN punches are available and recommended for creating clean, precise 
openings quickly and easily in most instrument panels.

 

If a punch is not available, a jigsaw or other cutting tool can be used. 
An adhesive template is provided to help guide the cutting process. 
De-burr and smooth the opening with a 

fi

 le.

1.  The panel mount transmitter is designed for installation using a ¼ DIN punch. 

Recommended clearance on all sides between instruments is 25 mm (1 in).

2.  Place gasket on instrument, and install in panel.
3.  Slide mounting bracket over back of instrument until bracket snaps into 

latches on sides of instrument.

To remove:

1.  Secure instrument temporarily with tape from front or grip from rear of 

instrument.
DO NOT RELEASE.

2.  Press bracket clips outward and remove.

Field Mount Installation

Field mounting requires a separate mounting kit. The 3-8050 Universal Mount Kit, the 
3-8051 or 3-8052 Integral Mount Kits, and the 3-9900.396 Angle Adjustment Adapter Kit 
enable the transmitter to be installed virtually anywhere. Detailed instructions for 

fi

 eld 

installation options are included with the 3-8050, 3-8051 and 3-8052 adapter kits 
(see Ordering Information section).

For Field Mount installations with a Cond/Res Module, the 3-9900.396 Angle Adjustment 
Adapter is required along with a 3-8050 or 3-8052 adapter kit to allow for suf

fi

 cient 

clearance for the wiring.

Field Mount with

3-8050 Universal Mount Kit

and Angle Adjustment Adapter

Field Mount with

3-8051 Integral Mount Kit

and Angle Adjustment Adapter

Содержание Signet 9900

Страница 1: ...085 0 01 cm 1 316 SS electrode Sanitary Tri clamp fl ange 1 to 1 in 3 2820 1 198 844 000 0 1 cm 1 316 SS electrode in threads in line only 3 2821 1 198 844 001 1 0 cm 1 316 SS electrode in threads in...

Страница 2: ...l warranty and non warranty repairs being returned must include a fully completed Service Form and goods must be returned to your local GF Sales office or distributor Product returned without a Servic...

Страница 3: ...the plug inmodules may be installed either before or after the base unit is mounted If the 9900 Base Unit will be mounted using the wall mount accessory kit 3 9900 392 install plug in modules first I...

Страница 4: ...s provided to help guide the cutting process De burr and smooth the opening with a file 1 The panel mount transmitter is designed for installation using a DIN punch Recommended clearance on all sides...

Страница 5: ...and mechanical damage Seal the cable entry points to prevent moisture damage Only one wire should be inserted into a terminal Splice double wires outside the terminal or use appropriate wire ferrule...

Страница 6: ...o be used for system power 10 8 to 35 2 VDC NOTE Backlight LEDs Modbus and optional Relay Module do not operate on loop power Any connected sensors or sensor electronics that cannot operate on loop po...

Страница 7: ...s Module which Provides filtering and signal conditioning The standard sensor cable length can be extended up to 30 m 100 ft Loop Voltage DC Power Cond Res Module SHIELD TEMP WHT ISO GND BLK SIGNAL RE...

Страница 8: ...PNP Style Wiring The 9900 Open Collector R1 output provides high speed switching capability Signal frequencies can reach 400 pulses per minute The 9900 Open Collector R1 output connection is dependen...

Страница 9: ...d is often required to start editing Once entered correctly this password will not be needed for subsequent edits However once the menu system is exited the password will again be required when edit m...

Страница 10: ...hanical Relay R2 Indicator LED Backlight Sensor do not block Warning LED Bar Graph Units of Measure GPM pH sec etc Value Label Menu Indication Menu Navigation Keys Mechanical Relay R3 Indicator LED Al...

Страница 11: ...new value selection 6 Press or to select another menu item Repeat steps 3 5 as required 7 Press again to return to normal operation The menu is constructed in a loop so you can move forward and backw...

Страница 12: ...display LOOKING FOR If no sensor is attached to the 9900 the words TYPE and COND RES are displayed If the 9900 does not identify your sensor type correctly use the and keys to select your sensor type...

Отзывы: