11
Signet 9900 Conductivity Transmitter
This basic operating procedure repeats
throughout the 9900 program:
1. Press ENTER for 3 seconds to enter
EDIT mode.
2. Press
►
to move to speci
fi
c menu
item.
3. Press ENTER key to select the item for
editing.
4. Press
▲
or
▼
to edit the value/
selection.
5. Press ENTER to store the new value/
selection.
6. Press
▲
or
▼
to select another menu
item. Repeat steps 3-5 as required.
7. Press
▲
+
▼
again to return to normal
operation.
The menu is constructed in a loop, so
you can move forward and backward to
select an item. After any item is saved
(by pressing ENTER), the display will
return to the previous menu.
System Setup: Menu Navigation
System Setup
ENTER
3s
Edit
Input Edit
Saves Changes
Input Edit Choices
(Password may
be required)
or
or
ENTER
or
ENTER
Select another
Menu Item
Return to
View Mode
+
2x
+
Or