Signet 2750 pH/ORP Sensor Electronics provide two-wire 4 to 20 mA loop output for 

pH and ORP measurements, without the expense of local display and other luxuries 

available in full-featured transmitters. Preampli

fi

 cation is built-in, reducing system costs 

while ensuring absolute signal integrity up to 305 m (1,000 ft).
These sophisticated 

fi

 eld-mount devices also provide digital (S

3

L) output. The 2750 

Sensor Electronics self-con

fi

 gure for pH or ORP operation via automatic recognition of 

electrode type, and the DryLoc

®

 electrode connector quickly forms a robust assembly 

for submersible and in-line installations.
NEMA 4X Junction Boxes are integral parts of the in-line versions (2750-1, 2750-2, 

and 2750-7) and are available as accessories for the submersible versions (2750-3 

and 2750-4). The optional EasyCal feature allows simple push-button calibration and 

includes an LED for visual feedback.

3-2750.090    Rev. N    10/16

*3-2750.090*

Signet 2750 DryLoc

®

 pH/ORP Sensor Electronics

English

Operating Instructions

Description

Table of Contents

2750-1, 2750-2

2750-3, 2750-4

Warranty Information ...............................................................................2
Product Registration ................................................................................2
Safety Information ...................................................................................2
Chemical Compatibility ............................................................................2
Speci

fi

 cations .......................................................................................2-3

In-Line Installation Notes .........................................................................3
In-Line Dimensions ..................................................................................3
Mounting Position ....................................................................................4
In-Line Assembly ..................................................................................4-5
Submersible Installation .......................................................................5-6
Digital (S

3

L) Wiring ..................................................................................6

4 to 20 mA Loop Wiring ...........................................................................7
Calibration ...............................................................................................7
Troubleshooting .......................................................................................8
Ordering Information................................................................................8

• 

English

• 

Deutsch

• 

Français

• 

Español

2750-7

Содержание Signet 2750-1

Страница 1: ...4X Junction Boxes are integral parts of the in line versions 2750 1 2750 2 and 2750 7 and are available as accessories for the submersible versions 2750 3 and 2750 4 The optional EasyCal feature allow...

Страница 2: ...00 or 3K Balco temperature sensor versions for pH Operational Range 2750 0 00 to 14 00 pH 2000 mV ORP 0 C to 85 C 32 F to 185 F Response Time includes electrode response pH 6 s for 95 of change ORP ap...

Страница 3: ...A Signet installation fitting 12 7 mm to 101 6 mm 1 2 in to 4 in or a threaded tee is required to secure the electrode in the pipe The Signet Measurement and Instrumentation catalog offers a complete...

Страница 4: ...nts or their vapor may lead to failure of the retaining nut ejection of the sensor and loss of the process fluid with possibly serious consequences such as damage to equipment and serious personal inj...

Страница 5: ...othing when working with chemicals in pressurized pipe Locked The molded alignment marks are enhanced for clarity 2750 In Line Assembly 1 Insert electrode into base of the 2750 3 or 2750 4 and turn un...

Страница 6: ...00 users If this direct wiring is used set the CALIBRATE menu to perform Calibration at INSTRUMENT Connect the terminals as shown to any digital S3 L terminals 8900 and 9900 users If using EasyCal set...

Страница 7: ...consequence For pH calibration use any two of these international standards pH 4 0 7 0 or 10 0 buffer solutions Signet part number 3 0700 390 contains one capsule of each value To produce standards f...

Страница 8: ...tor kit PG 13 5 1 connector 3822 7004 159 001 581 pH 4 01 buffer solution 1 pint 473 ml bottle 3822 7007 159 001 582 pH 7 00 buffer solution 1 pint 473 ml bottle 3822 7010 159 001 583 pH 10 00 buffer...

Отзывы: