Exiting Menu
Menu will repeat until
FLOW
is pressed.
4 to 20/0 to 20 mA (iso/
non-iso) options illustrated
MOD
Security
Code?
Yes
Unlock security
code using:
Select
next
item?
Yes
FLOW
Normal operation
ENTER
Recall original value
ENTER
Modify?
No
Yes
9 9 9 9
A N L 1 4 m
MOD
MOD
No
ENTER
Modify?
No
Yes
Select
next
item?
Yes
FLOW
9 9 9 9
A N L 1 2 0 m
MOD
Normal operation
Recall original value
MOD
ENTER
Modify?
No
Yes
Select
next
item?
Yes
Press to fine tune
4 mA output signal.
See note*
4
A D J U S T
MOD
FLOW
Normal operation
Recall original value
MOD
ENTER
Modify?
No
Yes
2 0
A D J U S T
MOD
MOD
Recall original value
OUTPUT
Press to select
flow rate at 20 mA
output signal
Press to select
flow rate at 4 mA
output signal
OUTPUT
OUTPUT
OUTPUT
Ammeter required,
see Figure 10
Press to fine tune
20 mA output signal.
See note*
Repeat
Menu?
Yes
FLOW
Normal operation
OUTPUT
Ammeter required,
see Figure 10
No
No
No
No
*Note:
Press:
FLOW
to
quickly access the minimum or
maximum current output signal
Press:
to fine
tune the current output signal
1 2 3 4 5 6 7 8 9 10 11 12 13 14
1 5 1 6 1 7 1 8 1 9 2 0 2 1 2 2 2 3 2 4 2 5 2 6 2 7 2 8
SS CNT GND PLS
OUTPUT
ANL2
R+ R-
SERIAL
Tx GND Rx
A in
A+ A-
BLK RED SHLD
SENSOR
L
1
N
AC IN
+ GND NO C NC
RLY1
DC IN
NO C NC
RLY2
R+ R-
ANL1
R+
R-
R+
R-
AMMETER
Figure 10
The min. and max. current
adjustment steps require an
external ammeter for monitor-
ing the current output.
3.6 Calibration Menu, Analog Outputs
17