Page 60 of 174
Troubleshooting
The table below describes the fault codes that may be generated and a possible action
for each fault code.
This may only be done by authorized personnel. The machine is connected to
the electricity supply and some components are live.
Fault
code
Fault
Comment
F00
Power failure
If the power failure lasts for more than 59 seconds, the process
is aborted. The fault code is generated when the power returns.
F01
The motor cutout has
tripped.
The motor cutout (-F01) for the circulation pump has tripped.
Possible action:
a. Check that there is power on all phases from the circulation
pump.
b. Check that the motor cutout is correctly set (see electrical
diagram).
c. Check that the pump impeller rotates easily.
d. Check the direction of rotation (see installation instructions).
F02
Water filling too slow.
The water level sensor (-B11) is not activated within two minutes
of the start of filling.
Possible action:
a. Check that the shutoff valves are open and that water is
reaching the machine.
b Check that the solenoid valves open and that their filters are
not clogged.
c. Check that the strainer in the machine is not blocked.
F03
Emptying time too long
Emptying takes longer than five minutes or the water level sensor
(-B11) is still activated when the emptying phase is complete.
Possible action:
a. Check that the waste pump (-M02) pumps out water.
b. Check that the pressure switch changes over.
c. Check that the solenoid valve closes (water does not leak in).
d. Check that the strainer in the machine is not blocked.
F04
Water leakage in the
dryer.
The switch to water level sensor (-B12) is open.
Possible action:
a. Check that the check valves to the drying connection close.
F05
Door fault, soiled side
Door does not lock (-S02) within 10 seconds or opens while a
program is running.
Possible action:
a. Check whether the microswitch (-S02) for the “door locked”
limit position has been activated within 10 seconds on door
locking.
b. Check that the motor (-M10) and the switch (-S52) for door
locking are working.
F06
Door fault, soiled side
The door does not open (-S01) within 10 seconds.
Possible action:
a. Check whether the microswitch (-S01) for the “door locked”
limit position has been activated within 10 seconds on door
locking.
b. Check that the motor (-M10) and the switch (-S51) for door
locking are working.
Содержание 86 SERIES
Страница 1: ...GETINGE 86 SERIES TECHNICAL MANUAL 502606900 SEV0725001...
Страница 10: ...Doc_TEC Doc_502606900 Rel_B Lang_GB Page 10 of 174 V1434 Steam heating Explanation of symbols V1560...
Страница 108: ...Doc_TEC Doc_502606900 Rel_B Lang_GB Page 108 of 174 List of components V1482...
Страница 126: ...Doc_TEC Doc_502606900 Rel_B Lang_GB Page 126 of 174...
Страница 127: ...Page 127 of 174 Doc_TEC Doc_502606900 Rel_B Lang_GB Electrical diagram...
Страница 128: ......
Страница 129: ......
Страница 130: ......
Страница 131: ......
Страница 132: ......
Страница 133: ......
Страница 134: ......
Страница 135: ......
Страница 136: ......
Страница 137: ......
Страница 138: ......
Страница 139: ......
Страница 140: ......
Страница 141: ......
Страница 142: ......
Страница 143: ......
Страница 144: ......
Страница 145: ......
Страница 146: ......
Страница 147: ......
Страница 148: ......
Страница 149: ......
Страница 150: ......
Страница 151: ......
Страница 152: ......
Страница 153: ......
Страница 154: ......
Страница 155: ......
Страница 156: ......
Страница 157: ......
Страница 158: ......
Страница 159: ......
Страница 160: ......
Страница 161: ......
Страница 162: ......
Страница 163: ......
Страница 164: ......
Страница 165: ......
Страница 166: ......
Страница 167: ......
Страница 168: ......
Страница 169: ......
Страница 170: ......
Страница 171: ......
Страница 172: ......
Страница 173: ......
Страница 174: ......
Страница 175: ......
Страница 176: ......
Страница 177: ......
Страница 178: ......
Страница 179: ......
Страница 180: ......
Страница 181: ......
Страница 182: ......
Страница 183: ......
Страница 184: ......
Страница 185: ......
Страница 186: ......
Страница 187: ......
Страница 188: ......
Страница 189: ......
Страница 190: ......
Страница 191: ......
Страница 192: ......
Страница 193: ......
Страница 194: ......
Страница 195: ......
Страница 196: ......
Страница 197: ......
Страница 198: ......
Страница 199: ......
Страница 200: ......
Страница 201: ......
Страница 202: ......
Страница 203: ......
Страница 204: ......
Страница 205: ......
Страница 206: ......
Страница 207: ......
Страница 208: ......
Страница 209: ......
Страница 210: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......
Страница 219: ......
Страница 220: ......
Страница 221: ......
Страница 222: ......
Страница 223: ......
Страница 224: ......
Страница 225: ......
Страница 226: ......
Страница 227: ......