Instructions for Use
Ambistik PFT-CPET
Serial no.: 2024xxxxx / 2025xxxxx
Version: 1 | Release date: 31 January 2022
Page 23
permitted by the manufacturer.
See chap. 12 "Technical Specifications"!
•
Do not connect any additional USB (except
mouse, keyboard, printer and other devices
which have not been approved by the
manufacturer!
•
Do not install any other software programs!
•
Ask your authorised specialist retail partner
which devices are approved by the manufactuer!
•
Ensure that in the patient environment – a
distance of 1.5 meters to the patient
−
there are
no accessible electrical parts (interfaces, plugs,
etc.) that are not isolated from the mains with an
isolation voltage of 4 kV!
•
Ensure that the users or third party persons do
not touch the patient and any conductive
connections or parts of the device that are
located outside the patient environment at the
same time!
Reason: Electric shock and / or misdiagnosis due to
loss of electrical safety caused by exceeding the
recommended maintenance schedule. Therefore:
•
Regularly check the specified maintenance
schedule!
•
If a maintenance schedule is exceeded, do not
continue to use the total system! Request
maintenance work from your authorised
specialist retail partner!
For this: Only use Ambistik PFT / Ambistik CPET within
its performance limits! Therefore:
•
Only use accessories, components and
consumable materials which have been
approved by the manufacturer, as well as
approved spare parts, optional add-ons and