Instructions for Use
Ambistik PFT-CPET
Serial no.: 2024xxxxx / 2025xxxxx
Version: 1 | Release date: 31 January 2022
Page 47
8
Servicing / Maintenance
8.1
Duties of the responsible Operator
Responsible organisations of medical devices are obligated, in
accordance with the corresponding, applicable regulations, to
preserve the properties of this medical device assured by the
manufacturer over its whole service life. This also includes
carrying out regular and proper servicing as well as safety
checks at intervals recommended by the manufacturer carried
out by specialist personnel who have been authorised by the
manufacturer for the respective tasks.
The expected service life of Ambistik PFT / Ambistik CPET is 8
years.
In its development, a great deal of value was placed on making
the servicing of all device components as simple as possible.
Only a little work is necessary to guarantee a fault-free operation
of the device.
However, to maintain a high quality of measurement results and
the safe operation, the manufacturer recommends having the
device safety and measurement precision checked by an
authorised specialist retail partner every 24 months.
These checks include:
•
Visual check
•
Electrical measurement
•
Functional Check
Possible danger to life.
Reason: Electric shock and / or misdiagnosis due to
loss of electrical safety caused by exceeding the
recommended maintenance schedule. Therefore:
•
Regularly check the specified maintenance
schedule!
•
If a maintenance schedule is exceeded, do not
continue to use the total system!
Request maintenance work from your authorised
specialist retail partner!