Instructions for Use
Diffustik / Variant PFTstik
Serial no.: xx|8|401|yyy and 2401xxxxx
Page 78
Version: 8 | Release date: 16. August 2021
The manufacturer accepts no liability for any resulting equipment
damage or any consequential damage caused by the use of
disinfectants that have not been tested and approved by the
manufacturer.
Attention
Diffustik resp. the variant PFTstik could be damage.
Reason: Improper treated components. Therefore:
•
For cleaning and disinfection only use those active
substances that are approved by the manufacturer
(see chap. 2.2.2 "Consumable Items / Auxilary
Materials")!
•
Follow the instructions on the concentration and dwell
time stated by the cleanser and disinfectant
manufacturer!
•
Do not put the flow unit and the shutter drive in
cleaning or disinfection fluid!
(Both are containing electronic components)
•
Do not disassemble the demand valve or put it into
disinfection fluid! Only surface disinfection!
Reason: Higher temperatures can lead to deformation and
damage. Therefore:
•
Observe the maximum permissible temperature of
75 °C for cleaning and disinfection!
•
Do not dry with heat!
Possible impairment of the measuring function.
Reason: Moisture in the connection port. Therefore:
•
Ensure that there is no moisture in the connection
piece before recommissioning!