Instructions for Use
Diffustik / Variant PFTstik
Serial no.: xx|8|401|yyy and 2401xxxxx
Version: 8 | Release date: 16. August 2021
Page 61
6.2.4
Power Supply and Computer Connection
Possible danger to life.
Reason: Electric shock due to lack of galvanic
separation with composition of non-approved
components. Therefore:
•
Only use the USB-Connection Cable
supplied by the Manufacturer as spare parts,
which are always part of the medical device!
•
The device may only be operated in conjunction
with the equipment cart (ergoline ergocar PC)
offered by manufacturer!
To make the Diffustik resp. the variant PFTstik ready for
operation, connect it:
1.
USB-connection cable
to USB socket of the Diffustik device
to free USB port of the computer
2.
Power supply cable
at socket Input AC 230 V and with
power cord to a socket
In addition for Diffustik resp. the
variant PFTstik with equipment cart
3.
Insert the power cord of the cart into
the socket
4.
Switch on the cart with On / Off
switch
Display “Power Supply“ of the
cart lights up