Instructions for Use
Diffustik / Variant PFTstik
Serial no.: xx|8|401|yyy and 2401xxxxx
Page 72
Version: 8 | Release date: 16. August 2021
Possible danger to life.
Reason: Electric shock. Misdiagnosis caused by
measurement error. For this: Ensure that the used
components are undamaged and that you are working
in a careful way! Therefore:
•
Prior to each use, visually inspect the total
system (housing, cables, connectors, tubings,
pneumatic connections, etc.) for any damages!
•
If there are any damages, do not operate the
system. The damaged parts must be replaced
or repaired properly!
Reason: Electric shock. Triggering malfunctions of
Diffustik resp. the variant PFTstik. Unqualified user /
operator can detect sources of failures too late or
even cause them. Therefore:
•
All maintenance and servicing work may only be
performed by specialised personnel who have
been authorised by the manufacturer!