This screen displays timer with elapsed seconds and EM61-MK2 readings. This display
lasts for 60 seconds. During this time please follow the instructions provided on the Allegro
screen.
Calibration can end in the following four ways:
The process can be stopped by an operator at any time by pressing key <Esc>.
The calibration may end by time out (60 seconds) if QC coil button was not
pressed or this action did not activate QC coil (see following Figures). After pressing
any key the program returns to Monitoring/Nulling manu. To repeat Internal QC
coil calibration please press key
I
again.
The logger will determine if the instrument has passed test. If after activating the
QC coil, the program determines that the reading is inside the standard values
/-10% tolerances then a message “CALIBRATION OK” will be displayed:
Otherwise, if the program determines that the reading is outside of the test values
range the corresponding message “CALIBRATION FAILED” is displayed.