2. QUICK START INSTRUCTIONS
For those familiar with Geotechnical instrumentation and its installation, the following quick
start instructions may be used. For more detailed instructions see Sections 3 and 4.
1)
Prior to installation, allow the piezometer to come to thermal equilibrium with the ambient
temperature for a minimum of 15 minutes. (Alternatively, if the instrument is attached to a
readout box, wait until the piezometer reading has stabilized.)
2)
Take an initial zero reading at zero (atmospheric) pressure.
3)
Verify that the initial zero reading is compatible with the factory supplied zero reading on the
calibration sheet.
4)
Record the barometric pressure and the temperature.
5)
Carefully measure and mark the cable where it will lie at the top of the borehole, well, or
standpipe, once the piezometer has reached the desired depth. (The piezo diaphragm lies 3/4
of an inch above the tip of the piezometer.)
6)
Saturate the piezometer filter. (See Section 3.1)
7)
For installation in standpipes or wells see Section 4.1, for boreholes Section 4.2, and for fills
and embankments Section 4.3.
3. PRIOR TO INSTALLATION
3.1 Saturating Filter Tips
Caution! - Do not allow the piezometer to freeze once it has been filled with water!
Most Geokon filter tips can be removed for saturation and then reassembled. To maintain
saturation, the unit should be kept underwater until installation. (
IMPORTANT
: The filter stone
and housing of model 4500C piezometers
are not removable
. Any attempt to remove the filter
stone or the housing will destroy the transducer!)
The saturation procedures are as follows:
3.1.1 Piezometers with low air entry (standard) filter housings
For accurate results, total saturation of the filter is necessary. For low air entry filters no
this saturation occurs as the tip is lowered into the water. Water is forced into the filter,
compressing the air in the space between the filter stone and the pressure sensitive
diaphragm. After a period of time, this air will dissolve into the water, filling the filter
and the space above it entirely with water.
To speed up the saturation process, remove the filter from the piezometer by carefully
twisting and pulling on the filter housing assembly (or unscrewing the point of the
piezometer for model 4500DP). Holding the piezometer so the filter is facing up, fill the
space above the diaphragm with water. Next, slowly replace the filter housing, allowing
the water to squeeze through the filter stone as it is installed. For piezometers with a
range of less than 10 psi, take readings with a readout box while reinstalling the filter
housing to ensure the piezometer is not overranged.