15
4.5 Model 4500H and Model 4500HH Transducer
Geokon Models 4500H and 4500HH are designed to be used in high temperature environments,
up to 250 degrees Celsius.
When connecting the Model 4500H transducer to external fittings, the fitting should be tightened
into the 1/4-18NPT female port by placing a wrench on the flats provided on the transducer
housing. Avoid tightening onto a closed system; the process of tightening the fittings could
overrange and permanently damage the transducer. If in doubt, attach the gage leads to a readout
box and take readings while tightening. For an easier and more positive connection to the
transducer, PTFE (plumber’s) tape on the threads is recommended. The maximum pressure for
the 4500H is 3MPa.
Geokon’s Model 4500HH is designed for high pressure environments. This model uses a 7/16-
20, 60 degree, female, medium pressure fitting. The maximum pressure for the 4500HH is
75MPa.
Both models employ a high temperature thermistor, see Appendix C.
CAUTION: All high pressure sensors are potentially dangerous. Care must be taken not to
overrange them beyond their calibrated range. Sensors are tested to 150% of their range to
provide a factor of safety.