28
Check the readout with another gage.
The Piezometer may have been overranged or shocked. Inspect the diaphragm and housing
for damage.
Symptom: Piezometer reading unstable
Connect the shield drain wire to the readout using the blue clip. (Green for the GK-401.)
Isolate the readout from the ground by placing it on a piece of wood or other insulator.
Check for sources of nearby noise such as motors, generators, antennas or electrical cables.
Move the piezometer cable if possible. Contact the factory for available filtering and
shielding equipment.
The Piezometer may have been damaged by overranging or shock. Inspect the diaphragm and
housing for damage.
The body of the Piezometer may be shorted to the shield. Check the resistance between the
shield drain wire and the Piezometer housing.
Symptom: Thermistor resistance is too high
Likely there is an open circuit. Check all connections, terminals, and plugs. If a cut is located
in the cable, splice according to instructions in Section 4.6.
Symptom: Thermistor resistance is too low
Likely there is a short. Check all connections, terminals, and plugs. If a short is located in the
cable, splice according to instructions in Section 4.6.
Water may have penetrated the interior of the piezometer. There is no remedial action.