10
Temperature scale
Celsius (°), Fahrenheit (°) oder Kelvin (K)
Temperature range
-50°C - 1300°C / -58°F - 2000°F /
223K - 2000K
Resolution
1°C or 1°F / 0,1°C or 0,1°F / 1K
Accuracy
-50°C - 0°C
± 2°C
0°C - 1000°C
± (0,5 % rdg + 1C°)
1000°C - 1300°C
± (0,8 % rdg + 1°C)
-58°F - 32°F
± 4°F
32°F - 2000°F
± (0,5 % rdg + 2°F)
223K - 273K
± 5K
273K - 2000K
± (1,0 % rdg +2 K)
Accuracy is specified for temperatures within the range of 18°C to
28°C (64°F to 82°F).
Temperature coefficient 0,1 times the applicable accuracy speci-
fication per °C from 0°C to 18°C and
28°C to 50°C (32°F to 64°C and 82°C to
122°C).
Input protection
60V DC or 24V AC
Reading rate
2,5 times per second
Input connector
Standard miniature thermocouple con-
nectors (flat blades spaced 7,9 mm centre
to centre)
Display
31/2 digits LCD with maximum reading
of “1999”
Battery
Standard 9V battery
Dimensions
162 × 76 × 39 mm
Weight
210 gr.
Supplied probe
4 foot type „K“ thermocouple bead probe
Length 1000 mm (teflon tape insulated)
Insulation temperature max. 260°C/500°F
Accuracy probe
±2,2°C or ±0,75 % or reading
(whichever is greater) from 0° -800°C
Working temperature
0°C - 50°C (32°F - 122°F)
Storage temperature
-20°C - 60°C (-4°F - 140°F)
Relative humidity
0 % - 80 % at 0°C - 35°C / 32°F - 95°F
Содержание FT 1300-1
Страница 23: ...23 ...