Symptom
Cause
Rectification
Samples not dry
IR lamp fault
See IR lamp / fan failure above
SampleGuard fault
See SampleGuard error above
Poor vacuum
Check vacuum connectors. Contact
Genevac service if leak cannot be
found
Aqueous samples freezing
due to evaporative cooling
Set controlled pressure to 8 mbar for
aqueous runs
Condenser over full
Defrost / drain condenser
Waste solvent container
valve closed
Open the waste solvent container valve
(see:
Waste solvent container
)
Condenser fails to drain
Possible failure of condenser drain
pump or blockage of drain pump filter.
Contact Genevac Service for
assistance.
Samples not dry on lower
rotor levels only
Lamp layers disabled
Go to
OPTIONS
menu and enable
lamp layers
Condensation inside
chamber
Chamber walls are too cold.
Refer to:
Condensation
Set chamber temperature control to
pre-heat chamber (specify
temperature)
Cross contamination
Samples bumping
Go to
Run Edit
screen and select the
Dri-Pure
option from the
Pressure
Control Regime
drop down menu
Set chamber temperature control to
wait to cool (specify temperature)
Pump runs in standby
Pump purge before power-
down
– HCl systems only
No fault, this is normal. Wait 45
minutes for pump to purge before
powering down system