Genevac HT-24
User Manual
16
10-1562 Issue 3-7 March
2014
For this reason it is advisable to position the channel 2 sample probe near the centre of the
holder as these samples can take longest to dry.
The thermal drain effect is particularly prevalent when using microtitre plates. Under these
conditions, dryness may occur in the outer wells whist the central samples remain in solution
or frozen. The use of Genevac Heat Transfer Plates reduces the effect by providing an even
heat transfer across the plate. For details of the range of heat transfer plates that are
available, contact your local Genevac representative to request a
Genevac Accessories
Brochure
or visit the Genevac website at
Selecting a run
From the
SELECT RUN
screen, use the
cursor direction keys
to move the white band up
or down the screen.
Cursor direction keys
The screen displays ten runs at a time and the next ten runs appear when the white band
scrolls off the top or bottom of the screen. Up to 100 runs can be stored in total.
Move the white band until it required run is highlighted.
.
Starting a run
With the required run highlighted on the
SELECT RUN
screen, press the
START
button to
start an evaporation using the selected run.
Note:
If the system is not ready, a message appears near the bottom of the screen, the run
will start once the system becomes ready.