Chapter 6 Application Settings
6.1 Application Setting Parameter table
parameter
Functional description
Incoming
and
outgoing
material
management
Minimum
incoming
weight
When the positive weight change is greater
than this value, the instrument considers
that feeding operation has occurred.If 0, no
judgment feeding operation is performed
Initial value: 1;Scope: 0-999999
Minimum
outgoing
weight
When the negative weight change is greater
than this value, the instrument considers
that discharging operation occurs;If 0, no
judgment operation will be performed
Initial value: 1;Scope: 0-999999
Feed in
and out
delay
In the set period of time, the weight does
not change, it is considered that the
material in and out of the end.
Initial value: 1.000s;Scope: 0.100 to
99.999
Automatic reporting
switch
After this function is enabled, the incoming
and outgoing material frames in GPRS
communication will be automatically
reported.Initial value: off;Range: on/off.
Application parameter
reset
Press ENTER (ENTER) and input the
calibration password to restore the factory
Settings of the calibration parameters.