Chapter 4 Setting parameters of weighing
parameter
Functional description
Reset
Settings
Automatic
zero
clearing
after
power-on
Set the reset range during power-on. If
the value is 0, the reset range will not
be reset during power-on.Initial value:
0;Parameter Range: 0 to 99%
Reset the
scope
Sets the range within which zeroing is
allowed.Initial value: 20%;Parameter
Range: 0 to 99%
STAB&TrZero
STAB
Timer
In this time, the weight variation range
is not greater than the range of
stability, considered stable.Initial
value: 1.000s;Parameter Range: 0.100 to
9.999 seconds
STAB
Range
During the stabilization time, the
weight variation range is not greater
than this set range, and it is
considered stable.Initial value:
1D;The value ranges from 0 to 99D
TrZero
Time
In the set time, the weight change range
is less than the zero range, automatic
tracking zero.Initial value:
1.000s;Parameter Range: 0.100 to 9.999
seconds
TrZero
Range
When the weight change range is less
than this set range during the zero
catching time, zero will be
automatically tracked.Initial value:
0d;The value ranges from 0 to 99D
Sampling
and
filtering
Digital
filtering
Digital filter, the higher the level,
the stronger the filtering, but the
longer the response time.Initial value:
4;Parameter Range: 0 to 9