Attach the arm on the OR-table rail; tighten the screw manually (39).
Adjust the swivelling rail clamp to the desired gradient and firmly tighten with the
locking handle (40).
IMPORTANT :
Check the tightness of the attachment before holding an instrument (grip on
the rail, fixation of the angular adjustment, and the overall stability and
rigidity of the rail itself).
Connect the supply hose (30) made from silicone with textile mesh to the “Stäubli”-
connector on the base below the central stand. The connector has a check valve
(31) and can be swivelled, however
do not remove the connector from the
central stand !
HOW TO FIX the swivelling rail clamp on the OR-table
III-2 Assemly of the instrument holder to the pneumatic central stand
First: Assembling the instrument holder:
Select the proper inlay for the tightening ball and insert it with the threaded end. Fix
it with the black ring on the opposite side. Turn the ring four times clockwise to fix
the inlay in the thread. Do not overtorque or tighten too much, so it will later allow
an easy insertion of the endoscope.
NOTE:
The tightening ball can be inserted or removed anytime once the lever is
pressed and the ball spanner (19) is disconnected.
UNCLAMPING
CLAM
PING