there are no conductive part of the instrument to come into contact with a
metal part of the arm (e.g., handle, part of an endoscope, etc.).
Unintended emission of substances or energy:
The Pneumatic Endoscope Holder is operated by pressurized gas. If the system
is not deflated by pushing the lever no gas is emitted.
Only in case the ball joints are released by pressure deflation, gas will be
emitting. The volume in case of simple repositioning under normal
atmospheric pressure is between 2 to 3 litres per minute.
When using a sterile cover the emitted gas will float to the rail clamp.
Energy emission:
No emission at all from electromagnetic energy (no power supply, consumption or
emission.
V - 2 Postioning of the Pneumatic Endoscope Holder
The easiest way to manipulating and using it is achieved by placing the Pneumatic
Central Stand on the rail opposite or lateral to the surgeon.
IMPORTANT: Avoid reducing or entirely obstructing the working space of the
surgical team.
V
– 3 sterile positioning with pneumatic central stand are unsterile
Not sterile pneumatic central stand
– sterile instrument holder