Field Calibration
4-7
September 2007
Two-Gas Field
Calibration
This simplified field calibration procedure uses two (zero and span)
gases to recalibrate the XMO2. Then, the XMO2 compares the data
from this field recalibration to the original factory calibration data,
and stores the difference as a
Drift Curve
.
Note:
If the range of your XMO2 is 0 to 21% O
2
, you can use air as
the span gas.
Setup
Before proceeding, you must be sure that your XMO2 is configured
for a two-gas calibration. The required reprogramming must be done
via the RS232 communication link, as follows:
1.
Enter the XMO2
Basic Menu
by slowly keying in
[Shift]+[1]
(if the
XMO2 is set up for
Easy Menu Entry
) or
[Enter], [1], [2], [3]
.
2.
One of the following initial menu options will appear:
3.
If the
Quick Offset?
option appears above, see Chapter 6,
General
Programming
, for instructions on reconfiguring the XMO2 for a
two-gas calibration.
Note:
The zero and span calibrations can be performed in either
order. For zero-based calibration ranges (e.g. 0-25%), we
recommend performing the span calibration first. For non-
zero-based calibration ranges (e.g. 90-100%), we recommend
performing the zero calibration first.
Proceed to the appropriate section to begin the field calibration.
Zero Gas Pushbutton
Calibration
To perform a zero gas pushbutton field calibration, complete the
following steps:
1.
Using the sample system controls, stop the flow of sample gas to
the XMO2 inlet port and initiate a flow of the same zero gas
specified on the XMO2 Calibration Sheet. Establish the same flow
rate and pressure conditions used for the sample gas, and allow the
zero gas to flow through the XMO2 for at least three minutes.
2.
Using Figure 4-2 on page 4-4 as a guide, locate the
Zero/Span
Selector
(Switch S1). Set the
Zero/Span Selector
(Switch S1) to
position “1" (“Zero”).
3.
Using Figure 4-2 on page 4-4 as a guide, locate the
Calibration
Pushbutton
(Switch S3). Depress the
Calibration Pushbutton
and
hold it down for 20 seconds.
BASIC MENU
Quick Offset? or Quick Zero?
Содержание XMO2
Страница 8: ...Chapter 1 ...
Страница 18: ...Chapter 2 ...
Страница 30: ...Chapter 3 ...
Страница 41: ...Chapter 4 ...
Страница 52: ...Chapter 5 ...
Страница 66: ...Chapter 6 ...
Страница 114: ...Chapter 7 ...
Страница 146: ...Chapter 8 ...
Страница 159: ...Appendix A ...
Страница 165: ...Appendix B ...
Страница 167: ...September 2007 Outline and Installation Drawings B 1 Figure B 1 Outline and Installation ref dwg 712 1008B ...
Страница 168: ...September 2007 Outline and Installation Drawings B 2 Figure B 2 XMO2 Assembly ref dwg 705 677H ...
Страница 171: ...September 2007 Outline and Installation Drawings B 5 Figure B 5 Digital PCB Assembly ref dwg 703 1139 ...
Страница 172: ...September 2007 Outline and Installation Drawings B 6 Figure B 6 Digital PCB Schematic ref dwg 700 1139 sht 1 ...
Страница 173: ...September 2007 Outline and Installation Drawings B 7 Figure B 7 Digital PCB Schematic ref dwg 700 1139 sht 2 ...
Страница 174: ...September 2007 Outline and Installation Drawings B 8 Figure B 8 Analog PCB Assembly ref dwg 703 1138 ...
Страница 175: ...September 2007 Outline and Installation Drawings B 9 Figure B 9 Analog PCB Schematic ref dwg 700 1138 sht 1 ...
Страница 176: ...September 2007 Outline and Installation Drawings B 10 Figure B 10 Analog PCB Schematic ref dwg 700 1138 sht 2 ...
Страница 177: ...Appendix C ...