Safety
2-24
Vscan Extend
–
User Manual
5721203-100
Rev. D
Maximum probe temperature
The table below indicates the maximum probe temperature.
NOTE:
Measurement uncertainty and probe variation: 2 °C.
NOTE:
Lens temperature measured under following conditions per EN/
IEC 60601-2-37:
•
Thermocouple was placed at the geometric center of the
lens.
•
Thermal phantom at 33 °C (or 23 °C) for external probes.
(Temperature rise is measured and added to 33 °C if the
phantom is at 23 °C).
NOTE:
Thermal phantom made with tissue-mimicking material as
referenced in EN/IEC 60601-2-37.
•
Probe placed upright in contact with above thermal
phantom.
•
Auto-freeze capability is disabled.
•
Lens temperature is monitored for 30 minutes.
Table 2-7:
Maximum probe temperature
Probe
Max Temp in deg C
(Simulated use)
Max Temp in deg C
(Still air)
Vscan Extend – configured with Sector probe:
Phased array transducer (G3S) for Deep scanning
39.7
43.1
Vscan Extend – configured with Dual probe:
Phased array transducer (G3S) for Deep scanning
41
42.6
Vscan Extend – configured with Dual probe:
Linear array transducer (G8L) allowing shallow
scanning
39.1
42.3
Содержание Vscan Extend
Страница 4: ...i 2 Vscan Extend User Manual 5721203 100 Rev D This page intentionally left blank ...
Страница 56: ...Safety 2 30 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 94: ...Preparing Vscan Extend for Use 3 38 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 212: ...Vscan Extend Maintenance 6 18 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 226: ...Appendix 7 14 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 248: ...Privacy and Security 8 22 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 251: ......