Vscan Extend Maintenance
6-14
Vscan Extend
–
User Manual
5721203-100
Rev. D
Troubleshooting
Vscan Extend troubleshooting
Table 6-2:
Troubleshooting
Problem
Possible cause
Solution
Vscan Extend has no power.
Battery not inserted.
Insert battery (see page 3-20)
Battery not charged.
Charge the battery (see
page 3-18)
Battery defect or end of life.
Contact GE Service (see
page 1-10)
Broken battery connection.
Contact GE Service (see
page 1-10)
Vscan Extend is not charging.
Battery not inserted.
Insert battery (see page 3-20)
Battery defect or end of life.
Contact GE Service (see
page 1-10)
Broken battery connection.
Contact GE Service (see
page 1-10)
Defective AC adapter or charger.
Contact GE Service (see
page 1-10)
Mains power is down.
Temperature is outside the
specified limits.
Ensure the ambient temperature
is within the specified limits (see
‘Environmental requirements for
the device’ on
)
Display screen is blank when the
device is powered on.
Connection broken during
software loading.
Contact GE Service (see
page 1-10)
Parts of the image is missing
when scanning.
Channels are missing.
Contact GE Service (see
page 1-10)
Noise when moving the probe
cable.
Defective probe cable.
Contact GE Service (see
page 1-10)
No image displayed when
scanning.
Defective probe.
Contact GE Service (see
page 1-10)
Содержание Vscan Extend
Страница 4: ...i 2 Vscan Extend User Manual 5721203 100 Rev D This page intentionally left blank ...
Страница 56: ...Safety 2 30 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 94: ...Preparing Vscan Extend for Use 3 38 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 212: ...Vscan Extend Maintenance 6 18 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 226: ...Appendix 7 14 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 248: ...Privacy and Security 8 22 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 251: ......