Cleaning agents
Follow the instructions described below while using CaviWipes
and Sani-Cloth HB.
CaviWipes
1. Use one CaviWipes towelette to completely preclean the
surface of the device, probe and cable.
2. Ensure that no debris and bioburden are left behind.
3. Discard
used
towelette.
4. Use a second CaviWipes towelette to thoroughly wet the
surface in order to disinfect the precleaned surface.
Repeated use of the product may be required to ensure that
the surface remains visibly wet for 3 minutes at room
temperature (20 degrees).
Sani-Cloth HB
1. Use one Sani-Cloth HB towelette to completely preclean the
surface of the device, probe and cable
2. Ensure that no debris and bioburden are left behind.
3. Discard
used
towelette.
4. Use a second Sani-Cloth HB towelette to thoroughly wet the
surface in order to disinfect the precleaned surface.
Repeated use of the product may be required to ensure that
the surface remains visibly wet for 10 minutes at room
temperature (20 degrees).
Содержание Vscan Extend
Страница 4: ...i 2 Vscan Extend User Manual 5721203 100 Rev D This page intentionally left blank ...
Страница 56: ...Safety 2 30 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 94: ...Preparing Vscan Extend for Use 3 38 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 212: ...Vscan Extend Maintenance 6 18 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 226: ...Appendix 7 14 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 248: ...Privacy and Security 8 22 Vscan Extend User Manual 5721203 100 Rev D ...
Страница 251: ......