Section 8-5 - Electronic Cage Components - Replacement Procedures
P R E L I M I N A R Y
2) Open and remove the Front End door assembly and place it on a stable surface:
3) Remove the CFE Board:
4.) Remove the ECG Board:
Removal of the PSB Board is now complete; front and rear views of the stand-alone component are
shown in
•
Front End Door Assembly Opening and Removal Procedure on page 8 - 83
•
CFE Board Removal Procedure on page 8 - 112
•
ECG/Respiratory Module Assembly Removal Procedure on page 8 - 106
Figure 8-146 PSB Board Removal Complete
Содержание Vivid S60N
Страница 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Страница 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Страница 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Страница 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Страница 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Страница 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Страница 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Страница 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Страница 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Страница 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Страница 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Страница 477: ......