D
IRECTION
FR091521, R
EVISION
1
V
IVID
S60N/V
IVID
S70N
B
ASIC
S
ERVICE
M
ANUAL
3-20
Section 3-5 - Completing the Setup
P R E L I M I N A R Y
The current drain will vary depending on the mains voltage.
•
At 230 VAC the current may be up to 2.2 A.
•
At 115 VAC the current may be up to 4.3 A.
•
At 110 VAC the current may be up to 4.6 A.
3-5-4
Connections on the I/O Rear Panel
NOTE:
Accessory equipment connected to the analog and digital interfaces must be certified according to the
respective IEC standards (e.g. IEC60950 for data processing equipment and IEC60601-1 for medical
equipment). Furthermore, all complete configurations shall comply with the valid version of the system
standard IEC60601-1-1. Everybody who connects additional equipment to the signal input part or signal
output part of Vivid™ S60/Vivid™ S70, configures a medical system, and is therefore responsible that
the Ultrasound system complies with the requirements of the valid version of IEC60601-1-1. If in doubt,
consult the technical service department or your local representative for GE.
3-5-4-1
Connect Ethernet
Connect the network cable to the Ethernet connector on the External I/O.
The connector is located on the rear side of Vivid™ S60/Vivid™ S70.
3-5-4-2
Connect USB Flash Card
NOTE:
USB Flash Cards approved for Vivid™ S60/Vivid™ S70 are verified for EMC performance according to
EN55011 class B. The use of any other USB Flash Cards will compromise this verification, and may
cause interference on Vivid™ S60/Vivid™ S70 itself, or on other electronic devices.
For approved models, please refer to
.
Insert the USB Flash Card in one of the USB ports on the Vivid™ S60/Vivid™ S70.
3-5-5
Connections on the Patient I/O Panel
The Patient I/O panel is located on the front of Vivid™ S60/Vivid™ S70.
Table 3-12 Electrical Specifications for all Vivid™ S60/Vivid™ S70 Models
Voltage
100-240 VAC
±10%
Power Consumption
500 VA
Frequency
50-60 Hz
Содержание Vivid S60N
Страница 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Страница 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Страница 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Страница 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Страница 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Страница 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Страница 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Страница 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Страница 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Страница 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Страница 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Страница 477: ......