Section 8-5 - Electronic Cage Components - Replacement Procedures
P R E L I M I N A R Y
8-5-22
Fan (AC Distribution Box) Replacement Procedure
8-5-22-1
Tools
Phillips screwdriver.
8-5-22-2
Preparation
Shut down the Vivid™ S60/Vivid™ S70 ultrasound unit, as described in
.
8-5-22-3
Time Required
10 minutes
8-5-22-4
Fan (AC Distribution Box) Removal Procedure
NOTE:
Access to the Fan is from the right side of the AC Distribution Box -
1.) Remove the AC Distribution Box and place it on a clean stable surface:
FRU Part #
Figure 8-172 AC Distribution Box Fan - Location
•
AC Distribution Box Removal Procedure on page 8 - 139
Fan
Содержание Vivid S60N
Страница 4: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ii ...
Страница 5: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iii ...
Страница 6: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL iv ...
Страница 7: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL v ...
Страница 8: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vi ...
Страница 9: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL vii ...
Страница 10: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL viii ...
Страница 11: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL ix ...
Страница 12: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL x ...
Страница 13: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xi ...
Страница 14: ...DIRECTION FR091521 REVISION 3 VIVID S60N VIVID S70N BASIC SERVICE MANUAL xii ...
Страница 477: ......