After The Exam Is Over
3-46
Versana Essential
–
User Guide
5759238-100 English
Rev. 2
Typical Tests to Perform
Quality assurance measurements provide results relating to
system performance. Typically these are:
• Axial
Measurement
Accuracy
•
Lateral Measurement Accuracy
•
Axial and Lateral Resolution
• Penetration
•
Functional & Contrast Resolution
•
Gray Scale Photography.
With these tests, a performance baseline can be set at
installation with the phantom in your department. Future test
results can be compared to the baseline in order to maintain a
record of system performance trends.
Frequency of tests
Quality assurance tests are used to determine whether a
scanner is providing the same level of performance from day to
day.
The frequency of testing varies with the amount of system usage
and modes to be tested. It is recommended that the user
perform quality assurance tests at least every three months or
every 400 patient studies. Tests should also be performed when
a question about system performance exists.
A mobile system may require more frequent tests.
Image quality should also be tested immediately after the
following events:
• Service
calls
• System
upgrades/modifications
•
Dropped probe, power surge, etc.