Measurement and Analysis
Versana Essential
–
User Guide
2-61
5759238-100 English
Rev. 2
Intensity (Echo level) measurement
To make an echo level measurement:
1. Press
the Measure
key in Measure mode to select
Intensity
.
2. To position the first caliper at the start point, move the
Trackball
.
3. To fix the trace start point, press
Set
. The caliper changes to
an active caliper.
4. To trace the measurement area, move the
Trackball
around
the anatomy. A dotted line shows the traced area.
5. To complete the measurement, press
Set
. The system
displays the echo level, as EL __ dB, in the Results Window.
NOTE:
The echo level measurement is only available on a frozen
image, not on a B-paused image.
NOTE:
Echo Level is not available through the factory default. To
enable echo level, modify the Measure Key Sequence preset,
found in the Utility -> Measure -> Advanced preset.