4°C to 40°C
Liquid temperature range
2 mm
Optical path length, 2 mm cell
10 mm
Optical path length, 10 mm cell
2 μL
Cell volume, 2 mm cell
8 μL
Cell volume, 10 mm cell
PTFE (polytetrafluoroethylene)
Wetted materials
PEEK (polyetheretherketone)
Titanium (palladium alloy)
Quartz (synthetic fused silica)
1–13, 13–14 (< 1 days exposure)
pH stability range
The wetted parts are resistant to organic solvents
and salt buffers commonly used in chromatogra-
phy of biomolecules, except 100% ethyl acetate,
100% hexane, and 100% tetrahydrofuran (THF)
or 15% THF in acetonitrile
Chemical resistance
UNF 10-32 fingertight connectors for tubing with
1/16" outer diameter
Tubing connections
UV Flow Cell ÄKTApilot 1/2/5 mm
800 ml/min
Recommended maximum flow
rate
2 MPa (20 bar, 290 psi)
Maximum pressure
Max. 0.05 bar at 400 ml/min with water at 25°C
Backpressure
4°C to 40°C
Liquid temperature range
1, 2 and 5 mm
Optical path length
300 μL
Cell volume
PEEK (polyetheretherketone)
Wetted materials
Titanium (palladium alloy)
FFKM (perflorelastoner) or EPDM
Quartz (synthetic fused silica)
1–13, 13–14 (< 1 days exposure)
pH stability range
Monitor UVis-920 Operating Instructions 29055049 AD
63
8 Reference information
8.1 Technical specifications