Display
Action
Step
Use the arrow keys to set the value. Use the fixed
values between 5.12 seconds and 0.08 seconds.
2
Min. peak width at half
height (s)
Time constant (s)
(approximate)
Averaging time (s)
32
2.0
5.12
16
1.0
2.56
8.0
0.5
1.28
3.2
0.2
0.64
1.6
0.1
0.32
0.8
0.05
0.16
0.5
0.03
0.08
Setting cell serial number, cell
type and cell path length
It is possible to store cell data in the instrument for up to five different cells. The data is
uniquely defined by each cells serial number.
Display
Action
Step
Select menu
Set Cell Serial Number
, press
OK
.
1
Enter the serial number of the cell (see the label on
the flow cell). Press
OK
. The serial number must be
1000 or higher.
2
Select menu
Set Cell Type
. Check the cell type value.
If the value is correct, use the
arrow
keys to go to
next menu. If the value has to be changed, press
OK
to change it.
3
Select menu
Set Cell Path Length
, press
OK
.
4
48
Monitor UVis-920 Operating Instructions 29055049 AD
5 Operation
5.8 Setup menu