Action
Step
Remove the four black protective caps from the optical fiber connectors.
2
Identify the fiber connector which has a gold colored knurled nut.
3
Attach this connector to the flow cell by carefully inserting it into the socket
and tightening the nut fingertight (fits only one of the two sockets).
4
Attach the other connector to the transmitter connector of Monitor UVis-920.
Do not overtighten.
Note:
The optical fiber with the gold colored knurled nut is the transmitter fiber and
must not be mixed up with the other fiber, the receiver fiber.
5
Connect the other optical fiber to the empty socket on the flow cell and to
the rear fiber connector on the monitor.
6
4.6
Connecting electrical cables
WARNING
Protective ground
. The product must always be connected to a
grounded power outlet.
WARNING
Power cord
. Only use power cords with approved plugs delivered
or approved by GE.
WARNING
Access to power switch and power cord with plug
. Do not block
access to the power switch and power cord. The power switch
must always be easy to access. The power cord with plug must
always be easy to disconnect.
The sockets for electrical signals are located on the rear panel. The instrument contains
no user replaceable fuse.
Monitor UVis-920 Operating Instructions 29055049 AD
33