Illustration
Action
Step
Fit the two small lugs on the cover in the holes at the
front and rear of the cell holder.
7
Lower the cover over the cell holder.
8
Installing UV Flow Cell ÄKTApilot
1/2/5 mm, and Industrial Flow
Cell 8 mm and 1"
Action
Step
Unpack the flow cell. Do not remove the black protective
caps on the optical fiber transmitter and receiver connector.
1
Choose the optical path length of the flow cell. The optical
path length can be changed with shim plates.
2
Attach the flow cell to an appropriate flow cell holder.
3
Connect the inlet and outlet tubing to the flow cell.
4
Connecting the optical fibers
WARNING
High intensity UV light
. This product uses high intensity ultra-violet
light. Do not disconnect the optical fibers while the lamp is on.
Monitor UVis-920 Operating Instructions 29055049 AD
31
4 Installation
4.5 Installing the flow cells