EQUIPMENT OVERVIEW: System Components
2-10
Solar 7000/8000/View Patient Monitor
Revision J
414993-001
What is a Tram-net
Hub Assembly?
If a patient monitor is connected to more than one peripheral device, the
Tram-net hub assembly is used. It connects the communication
processing capabilities inside the patient monitor to other equipment,
much like what a multiple outlet power strip does for ac power.
Peripherals can be connected to the Tram-net hub assembly via serial
cabling. The Tram-net hub assembly extends the patient monitor with a
cable of up to 1.3 meters (4 feet) long.
Note that a 25-pin D-type connector at the rear of the patient monitor is
marked TRAM-NET or ASYNC COMM. This connector handles both
Tram-net and async signals. One end of the Tram-net hub assembly
connects to this connector. At the other end of the Tram-net hub
assembly, the signals are separated into async and four Tram-net
connectors. The 25-pin red color coded connector handles async for
communication only with an async-only. The four 9-pin blue color coded
connectors are for extending the Tram-net network (blue label). More
details about Tram-net communication will be covered later in this
chapter.
Shown below is a Tram-net hub assembly. For details about the Tram-net
hub assembly, refer to the Modular Patient Monitor Accessories Service
Manual, pn 404183-150.
Содержание Solar 7000
Страница 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Страница 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Страница 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Страница 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Страница 20: ...EQUIPMENT OVERVIEW 2 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Страница 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 84: ...MAINTENANCE Checkout Procedure 3 22 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Страница 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Страница 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Страница 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 182: ...CONFIGURATION Advanced User Procedures 6 30 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Страница 184: ...UPPER LEVEL ASSEMBLY 7 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 213: ......
Страница 214: ......