CONFIGURATION: Advanced User Procedures
Revision J
Solar 7000/8000/View Patient Monitor
6-13
414993-001
Changing Software
Feature Level
The highest software feature level (i.e. 7015, 7020, 7025, 7030) of a
monitor is programmed into the serial EEPROM of the main processor
PCB. There are two ways of changing the software feature level. One way
is to dial a lower level and the other is to exchange the serial EEPROM
on the main processor PCB. The EEPROM should only be changed by
authorized personnel.
NOTE: The user may only change the feature level to a lower level than
the level installed at the factory.
Dial a Lower Level
There are times when a monitor may need to function at a lower feature
level than it was originally programmed at.
1.
To access FEATURE LEVEL option, execute the following menu
sequence starting from the MAIN menu:
MONITOR SETUP
SERVICE MODE
Password
Enter day and month from monitor screen with leading zeros.
(July 4 = 0407)
MENU SETUP
SOFTWARE LEVEL
2.
Rotate Trim Knob control to select the type of environment the
monitor will be used in.
3.
Press the Trim Knob control to exit. Your selection will be displayed
at the top of the screen after the time.
Exchange the EEPROM
To change the highest software feature level available in a monitor, you
must purchase the serial EEPROM U9 which contains the software
feature level you want.
Do the following to change the serial EEPROM in the monitor.
1.
Contact your Monitoring Sales Representative for the current part
number and order the part.
2.
Refer to the “Disassembly Guidelines” in Chapter 7:“Monitor
Assembly” of this manual.
Changing Remote
Control Function
There are three types of remote control available. In order to change the
function of a standard remote control to a neonatal or general operating
room (OR) remote control, a specific update kit and instructions are
required. Both kits listed below include the instructions, pn 404183-210.
Remote Control Update Kits
Item
Part Number
Operating Room Remote Control Update Kit
412977-002
Neonatal Remote Control Update Kit
412977-003
Содержание Solar 7000
Страница 1: ...Solar 7000 8000 SolarView Patient Monitor Field Service Manual 414993 001 Revision J ...
Страница 10: ...CONTENTS viii Solar 7000 8000 View Patient Monitor Revision H 414993 001 For your notes ...
Страница 11: ...Revision J Solar 7000 8000 View Patient Monitor 1 1 414993 001 1 INTRODUCTION ...
Страница 12: ...INTRODUCTION 1 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 19: ...Revision J Solar 7000 8000 View Patient Monitor 2 1 414993 001 2 EQUIPMENT OVERVIEW ...
Страница 20: ...EQUIPMENT OVERVIEW 2 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 63: ...Revision J Solar 7000 8000 View Patient Monitor 3 1 414993 001 3 MAINTENANCE ...
Страница 64: ...MAINTENANCE 3 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 84: ...MAINTENANCE Checkout Procedure 3 22 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 87: ...Revision J Solar 7000 8000 View Patient Monitor 4 1 414993 001 4 TROUBLESHOOTING ...
Страница 88: ...4 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 135: ...Revision J Solar 7000 8000 View Patient Monitor 5 1 414993 001 5 CALIBRATION ...
Страница 136: ...CALIBRATION 5 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 153: ...Revision J Solar 7000 8000 View Patient Monitor 6 1 414993 001 6 CONFIGURATION ...
Страница 154: ...6 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 182: ...CONFIGURATION Advanced User Procedures 6 30 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 183: ...Revision J Solar 7000 8000 View Patient Monitor 7 1 414993 001 7 UPPER LEVEL ASSEMBLY ...
Страница 184: ...UPPER LEVEL ASSEMBLY 7 2 Solar 7000 8000 View Patient Monitor Revision J 414993 001 For your notes ...
Страница 213: ......
Страница 214: ......