Using the Microsoft Windows App
78
SEER™ 1000
2067634-077 Revision B
If the connection cannot be established, you will be asked to try again.
NOTE
If the recorder is in the process of recording or contains a recording, a
warning is displayed.
Electrode Placement and Waveform Preview
WARNING
WAVEFORM DIAGNOSTIC QUALITY – The ECG waveforms shown on
the preview screen are used for assessing the quality of the connec-
tions only.
Do not use these preview waveforms for diagnostic purposes.
CAUTION
SIGNAL QUALITY – If there is a severe muscle interference or artifact
on the ECG when you tap the top of the electrodes lightly or when the
patient moves, it is possible that the electrodes are not attached
properly.
To increase the accuracy of analysis, make sure that the channel has
a high amplitude of QRS complex. If the amplitude is small, shift the
electrodes to find a suitable location for electrode attachment.
The app will display a torso image with the recommended electrode
placement. The system will automatically detect the cable type and dis-
play the appropriate configuration. Double-click the torso to zoom in or to
zoom out again.
NOTE
Click the
G
symbol to produce a screenshot of the waveform preview
screen for documentation purposes. This screenshot is placed into
the Windows clipboard.