9
2067634-077 Revision B
SEER™ 1000
93
9
Maintenance
Perform the following maintenance procedures as described.
Perform the visual inspection daily. If you notice any items that need re-
pair, contact an authorized GE Healthcare service person to make repairs.
Notices
NOTICE
DAMAGE TO THE EQUIPMENT – Some chemical substances can dam-
age the plastic of the device casing and cables. Also, some sub-
stances are conductive or insulating and can affect signal quality.
Use cleaning and disinfection agents only in accordance with the
manufacturer's instructions and, especially, be sure to use the
correct dilution factor given in these instructions.
Wring excess disinfectant from the wipe before using it.
Never immerse the device or cables in any liquid, as this may cor-
rode metal contacts and affect signal quality.
Do not allow fluid to collect around the connection pins. If this
happens, blot them dry with a soft, lint-free cloth.
Do not use solvents such as ether, acetone, or petroleum ether.
Never use conductive solutions or solutions that contain chlo-
rides, wax, or wax compounds to clean the device, cables, or
leadwires.