2026116-001 Revision B Responder™ 2000
Page 18
NOTES
NOTE:
Responder 2000, pads, and electrodes are latex-free.
NOTE:
Keep valid certificates of training and certification as required by state, province, or country regulations.
NOTE:
If the Battery Charger Charge Status light blinks red, a battery error has occurred during charging. If the Charge
Status light is solid red, a charger error has occurred during charging. Contact Customer Service in the event of an error
during charging.
NOTE:
If the Responder 2000 indicates an error code when powering on the device:
Do not use the Responder 2000 (Remove from patient)
Contact Customer Service with the error code(s).
NOTE:
If the system is pacing when the power button is pressed, a confirmation box displays requiring an additional press
of the Rotary Selector Knob before the system will turn off.
NOTE:
If the power button is pressed for five (5) seconds, the Responder 2000 will power down.
NOTE:
If AC power is not connected and the battery voltage becomes critically low, the system will display an error
message and then will power off.
NOTE:
If the ECG cable becomes disconnected or falls off, a warning message is displayed.
NOTE:
As the battery ages, it will discharge faster and there will be less operating time available before low battery
warning, therapy inhibit and system shutdown. Replace an aged battery to restore the operating time.
NOTE:
If the Responder 2000 issues an Error during this process:
Do not use the Responder 2000 (Remove from patient)
Write down any displayed error codes and contact Customer Service
NOTE:
Anterior / Posterior pads placement may alleviate a PADS SHORTED message.
NOTE:
The skin is a poor conductor of electricity, therefore preparation of the patient’s skin is important to facilitate good
electrode to skin contact. When cleaning the patient’s skin, NEVER use alcohol or tincture of benzoin, as this increases
skin resistance.
NOTE:
The selectable energy when using internal paddles is as follows: 2, 3, 5, 7, 10, 15, 20, 30, and 50 Joules.
NOTE:
Alarm Silence symbol is displayed, indicating that no audible tone warnings will be heard; only written warning
messages will be displayed on the graphics display.
NOTE:
The operator has thirty (30) seconds to deliver therapy before the Responder 2000 disarms and aborts therapy.
NOTE:
In Auto Sequence if the user wants to deliver a Sync Shock, Sync should be selected individually for all shocks in
Auto Sequence.
NOTE:
Every time after a Sync Shock is delivered; the device resets this toggle button to “No Sync”.
NOTE:
Verify printer has adequate paper on its roll for use.
NOTE:
To change the password, see TO CHANGE THE PASSWORD in this section of the manual.
NOTE:
All changes to the settings of the Responder 2000 must be performed before connecting the Responder 2000 to
the patient.
NOTE:
Changes to the power-up default in Menu, do not change on display.
NOTE:
The warranty will be void upon unauthorized disassembly or service of the Responder 2000.
Содержание Responder 2000
Страница 1: ...Operator s Manual 2026116 001 Revision B English ...
Страница 8: ...2026116 001 Revision B Responder 2000 Page 8 ...
Страница 46: ...2026116 001 Revision B Responder 2000 Page 46 ...
Страница 118: ...2026116 001 Revision B Responder 2000 Page 118 ...
Страница 141: ...2026116 001 Revision B Responder 2000 Page 141 ...
Страница 142: ...Section 7 Accessories 2026116 001 Revision B Responder 2000 Page 142 ...