2026116-001 Revision B Responder™ 2000
Page 72
The device defaults to the Demand mode and selects a pacer rate of 60 PPM (configurable).
1.
Select the Pacing Info Box by turning the Rotary Selector Knob.
2.
Turn pacer on.
3.
Select a low pacer current output, e.g. 20mA
4.
Select the required pacer rate.
5.
Now increase the pacer current output slowly until the heart reliably responds to the stimulation.
6.
Increase the pacer current output another 5 mA to ensure continued stimulation.
7.
Verify the stimulation by watching the ECG on the display.
8.
You can pause the stimulation and resume pacing with the same settings by turning focus to the “On” –field and
selecting “Pause” with the Rotary Selector Knob.
9.
After therapy, turn off the pacemaker before removing the pace pads carefully.