2026116-001 Revision B Responder™ 2000
Page 36
BUTTONS
There are 4 buttons on the Responder 2000:
1. Power
Button
2. Charge
Button
3. Shock
button
4. Manual
Button
POWER BUTTON
To Power on and off the Responder 2000, push the green Power button on the front panel of the Responder 2000.
POWER ON
1.
Press the green power button to turn on the Responder 2000. As the Responder 2000 powers on, the system
performs a self-test.
2.
After the Responder 2000 is powered on, it will automatically go into Manual Mode. The user can also program it to
enter Semi-Auto Mode or Monitor mode upon power on of the device.
NOTE:
If the Responder 2000 indicates an error code when powering on the device:
•
Do not use the Responder 2000 (Remove from patient)
•
Contact Customer Service with the error code(s).
POWER OFF
Press the green power button to turn off the Responder 2000.
NOTE:
If the system is pacing when the power button is pressed, a confirmation box displays requiring an
additional press of the Rotary Selector Knob before the system will turn off.
NOTE:
If the button is pressed for five (5) seconds, the Responder 2000 will power down.
NOTE:
If AC power is not connected and the battery voltage becomes critically low, the system will display an
error message and then will power off.
Содержание Responder 2000
Страница 1: ...Operator s Manual 2026116 001 Revision B English ...
Страница 8: ...2026116 001 Revision B Responder 2000 Page 8 ...
Страница 46: ...2026116 001 Revision B Responder 2000 Page 46 ...
Страница 118: ...2026116 001 Revision B Responder 2000 Page 118 ...
Страница 141: ...2026116 001 Revision B Responder 2000 Page 141 ...
Страница 142: ...Section 7 Accessories 2026116 001 Revision B Responder 2000 Page 142 ...