PHYSICAL DIMENSIONS
Height: 10.8 inches/27.4 cm
Width: 11.7 inches/29.7 cm
Depth: 7.4 inches/18.8 cm
Weight: Less than 10lbs/4.5kg, excluding battery, paddles, and a full roll of paper.
ENVIRONMENTAL REQUIREMENTS
OPERATING CONDITIONS
Temperature: 0
°
C to 50
°
C (32
°
F to 122
°
F)
Humidity: 10% to 95% RH, non-condensing
Air Pressure altitude: -500 ft (103kPa) to 15000 ft (57kPa)
STORAGE AND SHIPPING CONDITIONS
Temperature (Responder 2000): -20
°
C to 60
°
C (-4
°
F to 140
°
F)
Temperature (Pads): -12
°
C to 43
°
C (10
°
F to 110
°
F) for 2 years.
NOTE:
Do not exceed 38
°
C (100
°
F) for periods greater than 6 months in duration. Always store the pads in
their pouch to maintain freshness. Do not use if gel has dried out.
Humidity: 10% - 95% RH, non-condensing
Air Pressure altitude: -500 ft (103kPa) to 15000 ft (57kPa)
WARNING:
Electrode performance may be adversely affected by pre-attaching and storing with defibrillator cable or
exposure to air for long periods of time. These electrodes are not recommended for electrosurgery.
PRECAUTION:
Environment of use
Responder 2000 is designed for indoor use. Operator must confirm that the environment of use meets the required
operating environmental specifications before using.
PRECAUTION:
Cold Environments
If the Responder 2000 is stored in an environment with a temperature below the operating temperature, the unit should be
allowed to warm up to the needed operating temperature before using.
Содержание Responder 2000
Страница 1: ...Operator s Manual 2026116 001 Revision B English ...
Страница 8: ...2026116 001 Revision B Responder 2000 Page 8 ...
Страница 46: ...2026116 001 Revision B Responder 2000 Page 46 ...
Страница 118: ...2026116 001 Revision B Responder 2000 Page 118 ...
Страница 141: ...2026116 001 Revision B Responder 2000 Page 141 ...
Страница 142: ...Section 7 Accessories 2026116 001 Revision B Responder 2000 Page 142 ...