![GE PM880 Скачать руководство пользователя страница 44](http://html.mh-extra.com/html/ge/pm880/pm880_operations-and-installation-manual_128425044.webp)
Programming Site Data
3-5
August 2003
Entering Probe Configurations (cont.)
Note:
If you select “None,” you must enter constant values for
the PM880 to display any values.
a.
Press
[ENTER]
to open the drop-down menu.
b.
Use the [
T
] or [
S
] arrow keys to select the desired probe.
c.
Press
[ENTER]
to confirm the entry.
2.
If you choose to enter a constant for moisture, temperature or
pressure, you must enter a constant value. For a probe that
measures live inputs of a given parameter, you can enter a
optional constant value (as opposed to using the live
measurement). However, if the probe does not measure that
parameter (temperature or pressure), you cannot turn off the
constant value.
a.
Press the [
T
] arrow key to scroll to the “Use Constant
Value” section.
b.
If you can activate or deactivate a given value, the prompt
moves to the left-hand box. Press
[ENTER]
to turn the
value on or off. (For an activated constant, the box displays
a large “X,” while an empty box indicates a deactivated
constant.
c.
To modify the constant value, press the [
X
] key to move
to the text box, and press
[ENTER]
. Use the numeric keys
to enter the desired value, and press
[ENTER]
to confirm
the new value.
d.
Use the four arrow keys to move to other constant values,
and repeat steps b and c for any values you wish to
activate, deactivate or modify.
When you have completed entering probe and constant
information:
•
To confirm the entries and return to Operate Mode, press
[F3]
(OK). The PM880 returns to Operate Mode.
•
To leave the window without confirming the entries, press
[F2]
(Cancel) or the
[ESC]
key. The PM880 returns to Operate
Mode.
Содержание PM880
Страница 1: ...PM880 Portable Hygrometer ...
Страница 38: ...August 2003 2 18 Initial Setup Keypad cont Figure 2 8 The PM880 Keypad ...
Страница 150: ...August 2003 8 4 Servicing the PM880 Printing Reports cont Figure 8 3 Printout of a Typical Drive Report ...
Страница 237: ...5 August 2003 Index cont K K PPMv 3 17 Keypad Description 2 17 Key Description 2 17 Keys Description 2 17 ...