Diagnosis and Troubleshooting
9-3
August 2003
Table 9-1: Error Messages
Error Message
Problem
Action
Err 1: No Probe
A probe of type
required has not
been installed.
Connect the required
probe.
Err 2: Out of
Range
A calculated result
exceeds the unit’s
capacity.
Contact GE Panametrics.
Err 3: Over Range
A calculated result
exceeds the unit’s
calibration range.
Contact GE Panametrics.
Change the measurement
units so that the result is
within range.
Err 4: Under
Range
A calculated result is
below the unit’s cali-
bration range.
Contact GE Panametrics.
Err 5: Not Avail
The mode or units
selected require
more data or a dif-
ferent sensor.
Choose a different mode
and/or units.
Connect the required
probe.
Err 6: No Comm
The MIS Probe did
not respond to a
request for data.
Check the MIS Probe
connections.
Replace the MIS Probe.
Err 7: No Link
Communication
with an MIS Probe
has failed. The
probe is discon-
nected or damaged.
Check the MIS Probe
connections.
Replace the MIS Probe.
Err 8: Bad CRC
Communication
with an MIS Probe
is intermittent or dis-
torted.
Check for cable breaks or
high electromagnetic
interference (EMI).
Err 9: Bad Msg
The MIS Probe sent
a response that is not
supported by the
PM880.
Consult GE Panametrics.
Err 11: No Refer-
ence
No reference values
were entered for a
moisture or pressure
probe.
Enter reference values.