background image

20

EMC INFORMATION

EMISSIONS TEST & STANDARD 

COMPLIANCE

ELECTROMAGNETIC ENVIRONMENT - GUIDANCE

Radiated RF Emissions

CISPR 11

Group 1

The MRIaudio Sound System uses RF energy only for its internal 

function . Therefore, its RF emissions are very low and are not likely to 

cause any interference in nearby electronic equipment .

Conducted RF EMISSIONS

CISPR 11

Class A

The MRIaudio Sound System is suitable for use in all establishments 

other than domestic and those directly connected to the public 

low-voltage power supply network that supplies buildings used for 

domestic purposes .

Harmonic distortion

IEC 61000-3-2

Class A

Voltage fluctuations and flicker

IEC 61000-3-3

Complies

IMMUNITY TEST 

AND STANDARD

IEC 60601 TEST 

LEVEL

COMPLIANCE LEVEL

ELECTROMAGNETIC ENVIRONMENT - GUIDANCE

Electrostatic Discharge

IEC 61000-4-2

± 8 kV contact

± 2 kV, ±4 kV, ±8 kV, 

±15 kV air

± 8 kV contact

±15 kV air

Floors should be wood, concrete or ceramic tile . 

If floors are covered with synthetic material, the 

relative humidity should be at least 30% .

Voltage dips

IEC 61000-4-11

0 % UT; 0,5 cycle, @ 

0°, 45°, 90°, 135°, 180°, 

225°, 270° and 315°

0% UT; 1 cycle and 

70% UT; 25/30 cycles, 

@ 0°

0 % UT; 0,5 cycle, @ 

0°, 45°, 90°, 135°, 180°, 

225°, 270° and 315°

0% UT; 1 cycle and 

70% UT; 25/30 cycles, 

@ 0°

Main power quality should be that of a typical 

commercial or hospital environment . If the operator 

requires continued operation during power 

interruptions, it is recommended that the system be 

powered from an uninterrupted power supply or a 

battery .

Voltage interruptions

IEC 61000-4-11

0 % UT; 250/300 cycle

0 % UT; 250/300 cycle

Rated power freq . 

magnetic fields

IEC 61000-4-8

30 A/m

50 Hz or 60 Hz

30 A/m

50 Hz or 60 Hz

Power frequency magnetic fields should be at levels 

characteristic of a typical location in a typical 

commercial or hospital environment .

Electrical fast 

transients/burst

IEC 61000-4-4

± 2 kV for power

supply lines

± 1 kV for input/

output lines

100 kHz repetition 

frequency

± 2 kV for power

supply lines

± 1 kV for input/

output lines

100 kHz repetition 

frequency

Main power quality should be that of a typical 

commercial or hospital environment

Surges (Line-to-line)

IEC 61000-4-5

± 0 .5 kV, ±1 kV

± 1 kV

Main power quality should be that of a typical 

commercial or hospital environment .

Surges (Line-to-

ground)

IEC 61000-4-5

± 0 .5 kV, ±1 kV, ±2 kV

± 2 kV

The MRIaudio Sound System is intended for use in the electromagnetic environment specified below. The customer or the 

user of the MRIaudio Sound System should assure that it is used in such an environment .

Guidance and Manufacturer’s Declaration – Electromagnetic Immunity

U S E R   G U I D E

IM-1505 Rev. D

Содержание MRIaudio 1505

Страница 1: ...1 858 427 0679 support mriaudio com www mriaudio com Installation and User Guide Models 1505 and 1506 ...

Страница 2: ...t Numbers 7 INSTALLATION System Map Overview Model s 1505 and 1506 9 Cable Installation 10 Penetration Panel Connections 11 Sonic Transducer 12 Mounting Headphones 13 Digital Amplifier Assembly 14 Control Room Component Assembly 15 AutoVoice Adapter Connection 16 USER GUIDE iPad Interface Settings 17 Customer Support 17 How to Use the System 18 Troubleshooting 19 Frequently Asked Questions 19 SPEC...

Страница 3: ...he patient and or damage to this device please note the following warnings Check for any obvious damage prior to installing or using equipment If damage is detected refer to the standard warranty The wishbone In Ear Headset is much louder than the Over Ear Headphones Volume level should not exceed 40 Test the volume level prior to giving a patient the headphones Prolonged exposure to loud noises c...

Страница 4: ... equipment should be observed to verify the system is operating normally NOTE The emissions characteristics of this system equipment make it suitable for use in industrial areas and hospitals CISPR 11 class A If used in a residential environment for which CISPR 11 class B is normally required this equipment might not offer adequate protection to radio frequency communication services The technolog...

Страница 5: ...nonmagnetic non electrically conductive and non radio frequency reactive This eliminates all primary and potential risks during MRI scanning These objects are classified as MRI unsafe They are significantly ferromagnetic and pose a clear and direct threat to persons and equipment within the magnet room This system component may contain magnetic electrically conductive or radio frequency reactive p...

Страница 6: ...s 45 RF Shielded Cable Headphone Covers and Ear Tips optional 9 Pneumatic Tubing These system components are MR SAFE and can be used within the MRI bore during a scan WARNING These system components are MR CONDITIONAL and can be placed within the MRI Suite near or next to the bore These components should be kept at a minimum of 3 feet away S O U N D S Y S T E M C O M P O N E N T S IM 1505 Rev D ...

Страница 7: ...acement foam ear tips for In Ear Headphones Rated for 29 decibels noise attenuation Bio compatibility tested for patient safety In Ear Headphones 320 5759310 Slim wishbone design fits inside all MRI coils Provides 29 dB NRR Noise Reduction Rating when used with MRIaudio ear tips Over Ear Headphones 330 5759311 Audio headphone features over size ear cups that a provides 29 dB NRR Noise Reduction Ra...

Страница 8: ...ker Power Supply 670 20 5759319 Power supply for Technologist Speakers 5 8 cord type B US Canada Speaker Amplifier Cable 670 36 N A Connects Technologist Speakers to Digital Amplifier AutoVoice Adapter 25 pin 680 5759317 Integrates AutoVoice features with MRIaudio system Connects to control computer with Digital Amplifier 12 cable has 25 pin connector For GE MRI models Discovery AutoVoice Adapter ...

Страница 9: ...ded DB9 Cable 45 220 9 Pneumatic Tubing 9 600 10 Amplifier Power Supply 11 625 RCA to 3 5mm Cable 6 670 20 Technologist Speakers Power Supply 5 8 670 36 Technologist Speakers Amplifier Cable 3 680 or 685 AutoVoice Adapter 15 pin or 25 pin 12 The map diagram below provides an installation overview of the system that utilizes the 15 or 25 pin AutoVoice Adapter The system is designed within the expec...

Страница 10: ...gh the conduit to pull the new cable through Fig 2 Starting in the control room feed the fish tape through the conduit until it emerges in the computer room This will require periodic checking of the conduit in the computer room 1 4 Using electrical tape wrap the pigtail end of the new DB9 cable onto the end of the fish tape or the old existing cable Make sure to wrap it tightly and thoroughly in ...

Страница 11: ...e cable directly in front of the conduit pipe coil it then zip tie Leave enough slack to move the Digital Amplifier around the desk if necessary Stuff the zip tied cable up into the conduit and re fasten any access panels Penetration Panel A C A C A B B B A D C B Penetration Panel Penetration Panel Penetration Panel A B Penetration Panel A B Penetration Panel NEGATIVE J17 OR OTHER OPEN PORT POSITI...

Страница 12: ...er tabs from bottom of transducer and place where desired Feed Pneumatic Tubing through opening in machine base Fig 9 I N S TA L L AT I O N Fig 7 Fig 8 Fig 9 5 2 Connect the 45 cable to the BNC port on the transducer Fig 10 Pull any excess cable back toward the penetration panel leaving enough slack if the transducer needs to be adjusted Ensure that there are no coils in cable Coils are NOT MR SAF...

Страница 13: ...I machine approximately 4 5 feet off ground Fig 14 Remove the sticker backing on the In Ear Headphone hanger and attach under Over Ear mount Clamp the two headphones as shown Fig 15 6 2 To connect to music system plug the appropriate headphone set into Pneumatic Tubing Fig 16 Fig 14 Fig 16 Fig 15 IM 1505 Rev D ...

Страница 14: ...nce C Power Supply D 100 DB9 Cable E Bose Speakers 7 2 The Digital Amplifier will arrive pre configured DO NOT change the settings unless the system is not working correctly Check the back of the amplifier to ensure proper functionality ZONE 2 MIXER Dials 2 and 3 should be turned 75 clockwise Dial 1 should be turned fully counterclockwise NOTE Dials are plastic and should be turned with very littl...

Страница 15: ...INPUT 2 The white marked cord should go in white slot The red marked cord should go in red slot 8 2 Plug the Bose Speakers terminal clip 670 10 into the slot labeled ZONE 2 OUTPUTS 8 5 Plug the RCA cord audio jack into the left bottom audio port on the iPad 8 3 Plug the other end of the Bose Speakers connection cord into the slot labeled AUX IN Plug the power cord from the wall outlet into the mid...

Страница 16: ...4 Re connect the original pin connector into the back side of the AutoVoice Adapter Tighten all screws until snug Close panel and run cable up to Digital Amplifier 9 2 Using a small screwdriver detach the existing pin connector 9 3 Before installing run the AutoVoice Adapter cable through the opening in the computer panel Connect the new adapter back into the J7 port J7 IM 1505 Rev D 9 5 Plug Auto...

Страница 17: ... Settings FACETIME AND MESSAGES For any support questions or troubleshooting contact us through the FaceTime app for visual assistance A B C D iPad INTERFACE VOLUME POWER BUTTON SUPPORT SETTINGS RCA CORD AUDIO JACK LIGHTNING CHARGING PORT The iPad Mini arrives pre configured and is ready to use right away Slide the iPad into the locking mount and secure in place Lock combo is 0000 Ensure that the ...

Страница 18: ...e head band for size and place onto patient Ensure the ear cups are covering entire ear For In Ear Headphones Gently insert foam tips into patient s ears covering the entire ear canal 10 8 MIC AND AUDIO CONTROL KNOBS The knob MIC INPUT 1 controls the technologist microphone to the patient The knob MUSIC INPUT 2 controls the patient s headphone volume Both knobs have a green indicator light when in...

Страница 19: ...rating problems Does MRIaudio create noise or artifacts The MRIaudio system is designed from the ground up to eliminate the likelihood of artifacts or noise through the use of specialized components made specifically for compatibility in the MRI environment This has been proven through rigorous third party testing and vetted by the U S Food and Drug Administration Are MRIaudio systems OEM compatib...

Страница 20: ... 1 cycle and 70 UT 25 30 cycles 0 Main power quality should be that of a typical commercial or hospital environment If the operator requires continued operation during power interruptions it is recommended that the system be powered from an uninterrupted power supply or a battery Voltage interruptions IEC 61000 4 11 0 UT 250 300 cycle 0 UT 250 300 cycle Rated power freq magnetic fields IEC 61000 4...

Страница 21: ...rom structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measure...

Страница 22: ...moving company that always aims to go above and beyond heeding the voice of the customer at every step Working directly alongside MRI technologists MRIaudio consistently utilizes their feedback to improve our products as evidenced by the suite of customer suggested features that have been included in the third generation of the MRIaudio system Our president and CEO Spencer Howe believes that simpl...

Отзывы: